Technical Memorandum 17 October 1993
Revision date: October 2010
555 North 30th Street, Omaha, NE 68131
(402) 498-6700
BOYS TOWN NATIONAL RESEARCH HOSPITAL
PUTT User Manual
Stephen Neely
ABSTRACT
PUTT is a program for measuring and displaying hearing thresholds. The
name PUTT is derived from “pure-tone thresholds.” It was designed to
investigate whether sound intensity level (SIL) could eliminate the
problem of standing waves when sound levels are measured in the ear
canal. Recent versions include forward pressure level (FPL) as
alternative way to quantify sound level. PUTT implements both maximum
likelihood and staircase tracking methods to determine threshold. Tones
are presented in a single interval, yes/no paradigm, or in a two-interval
forced choice paradigm. Parameters of the stimulus presentation and
threshold determination are specified in a configuration file.
PUTT will
work with most high-quality soundcards (e.g., CardDeluxe, Gina3G,
Indigo io) and uses synchronous averaging for stimulus calibration.
This document describes PUTT version 2.29.
Page view 0
1 2 3 4 5 6 ... 16 17

Summary of Contents

Page 1 - PUTT User Manual

Technical Memorandum 17 October 1993 Revision date: October 2010 555 North 30th Street, Omaha, NE 68131 (402) 498-6700 BOYS TOWN NATIONAL RESEARCH

Page 4 - 3. Configuration parameters

APPENDIX A: Program Installation Windows. The most recent version of PUTT may be downloaded at http://audres.org/downloads/putt-setup.zip. Inside t

Page 5

APPENDIX B: SOUNDCARD INSTALLATION CardDeluxe The CardDeluxe is manufactured by Digital Audio Labs. http://www.digitalaudio.com/ Connection of the Ca

Page 6

APPENDIX C: CAVITY CONSTRUCTION The probe calibration cavities are constructed from brass tubing in the dimensions outlined below. A tube cutter is

Page 7 - 4. Screenshots

APPENDIX D: CALIBRATION FILES When PUTT performs in-the-ear calibration, the stimulus waveform and two microphone responses (one for each loudspeaker

Page 8

px nf+1 complex double crosstalk pressure (Pa/V) df 1 double frequency bin (Hz) dc 1 double cavity diameter (cm) tc 1 double cavity temperatu

Page 9

Page 17 INDEX CardDeluxe ... 10 emav.ini ...

Page 10 - Page 10

Table of Contents 1. Introduction ...

Page 11

1. Introduction PUTT is a contraction of “pure-tone thresholds.” A special feature of the program is its ability to quantify sound levels in terms o

Page 12

second interval. The threshold sound levels for selected set of frequencies is stored in a TTH file that has the same name as CAL file that contains t

Page 13

DA2_sensitivity = 0 Sensitivity of the second (or Right) D/A converter in count per volt. When set to zero, the value is calculated from ARSC registry

Page 14

Freq_start = 500 Starting frequency (Hz). Freq_stop = 40 Highest frequency (Hz). Per_octave = 1 Number of frequencies per octave. Tone_duration = 0

Page 15

Protocol = 1 Frequency set (1=“standard”, 2=“F1, F2, Fd”) Scope = 2048 Number of waveform points displayed during calibration. hfthr = N Add 9 &